Deprecated: substr(): Passing null to parameter #1 ($string) of type string is deprecated in /var/www/html/tmp/cache/skel/html_8eb2024d66cfd26a3bedd5c1f80f9d34.php on line 25
TY - JOUR
TI - Design and Invitro Studies of a Needle-Type Glucose Sensor for Subcutaneous Monitoring
AU - Bindra, D.S.
AU - Zhang, Y.N.
AU - Wilson, G.S.
AU - Sternberg, R.
AU - Thevenot, D.R.
AU - Moatti, D.
AU - Reach, G.
T2 - Analytical Chemistry
AB - A new miniaturized glucose oxidase based needle-type glucose microsensor has been developed for subcutaneous glucose monitoring. The sensor is equivalent in shape and size to a 26-gauge needle (0.45-mm o.d.) and can be implanted with ease without any incision. The novel configuration greatly facilitates the deposition of enzyme and polymer films so that sensors with characteristics suitable for in vivo use (upper limit of linear range > 15 mM, response time < 5 min, and sensitivity yielding a 5:1 signal-to-background ratio at normal basal glucose levels) can be prepared in high yield (> 60%). The sensor response is largely independent of oxygen tension in the normal physiological range. It also exhibits good selectivity against common interferences except for the exogenous drug acetaminophen.
DA - 1991///
PY - 1991
VL - 63
IS - 17
SP - 1692
EP - 1696
ER -